Air Cooled Thermoelectric Cold Plates For Laboratory
The cold plates on this page are designed to cool and heat, through direct contact, laboratory samples or other loads. Reaching temperatures as cold as -50C is possible with the Cascade system. RS-232 and other features are included ... read more ➡ |
|
AHP-1800CPV Series
Style: Bench Top
Voltages: 100 - 240 VAC
Environments: Bench Top, Laboratory, Industrial
|
Cooling capacity:
400 - 430 Watts @ 0 °C
ΔT
Includes:
Integral Power Supply
Integral PWM Temperature control
USB Communications
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
NOTES |
COLD PLATE |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT °C |
AHP-1800CPV |
9-04KB-1-0A0 |
Cool/Heat |
Smooth Surface |
100-240 |
5.0-2.5 |
0-40 |
AHP-1800CPV |
9-04KB-1-TAP |
Cool/Heat |
6-32 Tap Pattern |
100-240 |
5.0-2.5 |
0-40 |
AHP-1800CPV |
9-04KB-1-MET |
Cool/Heat |
M3 Tap Pattern |
100-240 |
5.0-2.5 |
0-40 |
Cold Plate Features:
•Precision machined cold plate surface 13.3" X 12.8"
•Cools and heats (-15 °C to 90 °C)
•Easy clean stainless steel top surface
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Variable speed fans for quieter operation
•Compact bench top unit,18.5” X 15.1” footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Many options and accessories available |
| Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
Click on the image for a larger view
Click on the image for larger view of the performance curve
|
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the images to open the PDF files
|
|
|
AHP-1200DCP Series
Dual Temperature Zone Plate
Style: Bench Top
Voltages: 100-240 VAC
Environments: Bench Top, Laboratory, Industrial |
Cooling capacity:
400 - 480 Watts @ 0 °C
ΔT (combined)
200 - 250 Watts @ 0 °C
ΔT (individual plate)
Includes:
Integral Power Supply
Two integral PWM temperature controllers
Two USB comms ports
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
NOTES |
COLD PLATE |
VOLTAGE VAC |
CURRENT AMPS. |
OPERATING AMBIENT °C |
AHP-1200DCP |
9-34KB-1-0A1 |
Heat/Cool |
Smooth Surface |
100-240 |
2.5-5.0 |
0-40 |
AHP-1200DCP |
9-34KB-1-TA1 |
Heat/Cool |
6-32 Tap Pattern |
100-240 |
2.5-5.0 |
0-40 |
AHP-1200DCP |
9-34KB-1-ME1 |
Heat/Cool |
M3 Smooth Surface |
100-240 |
2.5-5.0 |
0-40 |
Cold Plate Features:
• Cools and heats two plates independently
• 100-240 VAC universal input
• Low-profile design with ergonomic sloped front
• Quiet operation
• Compact bench top unit,18.5”X 15.1”footprint
• No compressor,fluorocarbons or filters.
• Virtually maintenance-free operation
• Painted Enameled stainless steel exterior housing
• Rubber feet
• Precision machined cold plate surface
• Large 13.5" x 13" overall plate area
• Adaptable to various surfaces/coverings
• Use with clear plastic walls or covers |
Control Features:
• Integral "tunable" PWM temperature control
• Bi-directional Temperature Control for Thermoelectric applications
• 4 Temperature zone with 4 independent PID settings
• Heating and Cooling •Internal RTD sensor
• Remote Sensibility™ switchable exterior sensor
• Output PWM 1KHz, duty cycle resolution=1000 steps
• Labview VI examples available
• USB communications |
Applications:
• Behavior Studies• Habitat studies• Pain Threshold Studies• Temperature Range Studies• Heat/Cold Sensitivity• Temperature Differential Testing• Long Term Temperature Exposure• Heat/Cold Discrimination• Heat/Cold Transient Studies• Specimen Storage/Temperature Maintenance• Histology Sample Preparation• General Laboratory Cold Plate Use |
Click on the image for a larger view
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to open the PDF fil
|
|
|
AHP-1200CPV Series
Style: Bench Top
Voltages: 100 - 240 VAC, 24 VDC
Cold Plate Temperature Range: -20 °C to 90
°C
Environments: Bench Top, Laboratory, Industrial |
Cooling capacity:
230 - 240 Watts @ 0 °C
ΔT
Includes:
Integral Power Supply
Integral PWM Temperature control
USB comms port
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
NOTES |
COLD PLATE |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT °C |
AHP-1200CPV |
9-35KB-1-0A0 |
High capacity |
Smooth Surface |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35KB-1-TAP |
High capacity |
6-32 Tap Pattern |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35KB-1-MET |
High capacity |
M3 Tap Pattern |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35KB-1-TAP |
High capacity |
6-32 Tap Pattern |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35K5-1-TAP |
High capacity |
6-32 Tap Pattern |
24 VDC |
17 |
0-50 |
AHP-1200CPV |
9-35K5-1-MET |
High capacity |
M3 Tap Pattern |
24 VDC |
17 |
0-50 |
AHP-1200CAS* |
9-35KB-1-CAS* |
High capacity |
Tap Pattern |
100-240 |
4.0† |
0-50 |
This part number is ready for use with a low temperature cascade option and includes CH-1200 hinged cover,CC-1200 rear panel for cascadepower up and control,refer to pages 80 and 81 for information on cascades and other available options.
† Reflects the current draw of the unit @ 120 VAC, 60 Hz input.
Cold Plate Features:
•Precision machined cold plate surface
•Easy clean stainless steel top surface
•Cools and heats (TYP.-20 °C to 90 °C)
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Variable fan speed for quieter operation
•Weighs only 25 lbs.(11.4 kg)
•Compact bench top unit,11.2”X 15.1”footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Accessories for glassware (beaker/test tube) cooling
•Many options and accessories available |
Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
|
Click on the images for larger view
Shown above AHP-1200CPV and AHP-1200CPV with CH-1200 hinged cover accessory.
Click on the image for a larger view
View or request 3D solid model files for this product:
Clic on the image to open the PDF file
|
|
|
AHP-1200CPV High Temp. Series
Style: Bench Top
Voltages: 100 - 240 VAC, 24 VDC
Cold Plate Temperature Range: -20 °C to 120
°C
Environments: Bench Top, Laboratory, Industrial |
Cooling capacity:
230 - 240 Watts @ 0 °C
ΔT
Includes:
Integral Power Supply
Integral PWM Temperature control
USB comms port
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
NOTES |
COLD PLATE |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT °C |
AHP-1200CPV |
9-35KB-6-0A0 |
High capacity |
Smooth Surface |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35K5-6--0A0 |
High capacity |
Smooth Surface |
24 VDC |
17 |
0-50 |
AHP-1200CPV |
9-35KB-6-TAP |
High capacity |
6-32 Tap Pattern |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35KB-6-MET |
High capacity |
M3 Tap Pattern |
100-240 |
3.5† |
0-50 |
AHP-1200CPV |
9-35K5-6-TAP |
High capacity |
6-32 Tap Pattern |
24 VDC |
17 |
0-50 |
AHP-1200CPV |
9-35K5-6-MET |
High capacity |
M3 Tap Pattern |
24 VDC |
17 |
0-50 |
AHP-1200CAS* |
9-35KB-6-CAS* |
High capacity |
Tap Pattern |
100-240 |
4.0† |
0-50 |
* This part number is ready for use with a low temperature cascade option and includes CH-1200 hinged cover,CC-1200 rear panel for cascadepower up and control,refer to pages 80 and 81 for information on cascades and other available options.
† Reflects the current draw of the unit @ 120 VAC, 60 Hz input.
Cold Plate Features:
•Precision machined cold plate surface
•Easy clean stainless steel top surface
•Cools and heats (TYP.-20 °C to 120 °C)
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Variable fan speed for quieter operation
•Weighs only 25 lbs.(11.4 kg)
•Compact bench top unit,11.2”X 15.1”footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Accessories for glassware (beaker/test tube) cooling
•Many options and accessories available |
Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
|
Click on the image for a larger view
Shown above AHP-1200CPV and AHP-1200CPV with CH-1200 hinged cover accessory.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the imge to open the PDF file
|
|
|
TGB-5030 Thermal Gradient Bar
Gradient Area: 500 X 300 mm
Material: ALuminum, clear anodized
Embedded Sensor: 3 Wire RTD, 100 ohm |
TECAs TGB-5030 Gradient Bar option expands the use of the AHP-1200CPV line of Cold Plates. With TECA model AHP-1200CPV cold plate at each end the gradient bar has programmable gradient areas with near linear gradients between the two ends and near uniform temperatures along the width of the bar. The temperature at each end of the gradient area can be programmed to a single set point or with a ramp/step and soak profile creating constant, expanding, contracting or moving gradients. Set the gradient profile to be large for initial observations then adjust the end set points to zoom in on a temperature range of interest. The temperature at the end of the gradient area is measured from underneath the plate at approximately .05 from the surface leaving the top gradient area clean and smooth. Optional external surface sensors can be used to change the size of the gradient area. |
Click on the image for a larger view
The picture on the right shows TGB-2515 Thermal Gradient Bar used with two AHP-301CPV and the picture on the left shows TGB-5030 Thermal Gradient Bar used with two AHP-1200CPV.
Click on the image for a larger view of the performance curve
TGB-2515 Thermal Gradient Bar
|
TGB-5030 Thermal Gradient Bar |
Click on the image for a larger view
Click on the image to view the PDF file
|
|
|
AHP-1200CAS Cascade Series
Style: Bench Top
Voltages: 100 - 240 VAC, 24 VDC
Cold Plate Temperature Range: -50 °C to 90
°C
Environments: Bench Top, Laboratory, Industrial |
Maximum ΔT @ 0 Watts Load:
-45 °C for AHP-1200CAS
-59 °C for AHP-1200C22
-68 °C for AHP-1200C21
-76 °C for AHP-1200C31
Referenced 25 °C operating ambient
Includes:
Integral Power Supply
Integral PWM Temperature control
USB comms port
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
INCLUDED
CASCADE |
COLD PLATE
SURFACE in |
MAXIMUM
delta T °C |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT RANGE °C |
COLD PLATE TEMPERATURE RANGE °C |
AHP-1200CAS |
9-35KB-1-CAS |
None (base model) |
13.3X5.77 |
-45 |
100-240 |
6.2 |
0/50 |
-20/90 |
AHP-1200C22 |
9-35KB-1-C22 |
CCP-22 |
8.75X4.50 |
-59 |
100-240 |
6.6 |
0/50 |
-30/90 |
AHP-1200C21 |
9-35KB-1-C21 |
CCP-21 |
6.0X3.0 |
-68 |
100-240 |
6.4 |
0/50 |
-40/90 |
AHP-1200C31 |
9-35KB-1-C31 |
CCP-31 |
6.0X3.0 |
-76 |
100-240 |
6.8 |
0/50 |
-50/90 |
The current draw of the unit reflect the operation in 120 VAC 60 Hz.
Cold Plate Features:
•Precision machined cold plate surface
•Easy clean stainless steel top surface
•Cools and heats
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Variable fan speed for quieter operation
•Weighs only 25 lbs.(11.4 kg)
•Compact bench top unit,11.2”X 15.1”footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Accessories for glassware (beaker/test tube) cooling
•Many options and accessories available |
Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
Click on the image for a larger view
Click on the image for larger view of the performance curve
|
Other test curves for AHP-1200CAS which might be of interest:
- CCP-22 with AHP-1200CPV CAS, 60 °C to -30 °C cooling rate, CCP-22 was covered
- CCP-22 with AHP-1200CPV CAV, -35 °C to -20 °C step change, CCP-22 was covered
- CCP-31 with AHP-1200CPV CAS range
- CCP-31 with AHP-1200CPV CAS, 1st half insulated and covered, 2nd half covered only
- CCP-31 with AHP-1200CPV CAS response time,
1st half insulated and covered, 2md half covered only
- CCP-31 with AHP-1200CPV CAS response time fan speed @ 50%, covered only
- CCP-31 with AHP-1200CPV CAS delta T, fan speed @ 50%, covered
|
|
|
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view AHP-1200CAS 3D PDF file: |
|
Click on the image to open the PDF file
|
|
AHP-1200CAS Cascade High Temp. Series
Style: Bench Top
Voltages: 100 - 240 VAC, 24 VDC
Cold Plate Temperature Range: -50 °C to 150
°C
Environments: Bench Top, Laboratory, Industrial |
Maximum ΔT @ 0 Watts Load:
-45 °C for AHP-1200CAS
-59 °C for AHP-1200C22
-68 °C for AHP-1200C21
-76 °C for AHP-1200C31
Referenced 25 °C operating ambient
Includes:
Integral Power Supply
Integral PWM Temperature control
USB comms port
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
INCLUDED
CASCADE |
COLD PLATE
SURFACE in |
MAXIMUM
delta T °C |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT RANGE °C |
COLD PLATE TEMPERATURE RANGE °C |
AHP-1200CAS |
9-35KB-6-CAS |
None (base model) |
13.3X5.77 |
-45 |
100-240 |
6.2 |
0/50 |
-20/120 |
AHP-1200C22 |
9-35KB-6-C22 |
CCP-22H |
8.75X4.50 |
-59 |
100-240 |
6.6 |
0/50 |
-30/150 |
AHP-1200C21 |
9-35KB-6-C21 |
CCP-21H |
6.0X3.0 |
-68 |
100-240 |
6.4 |
0/50 |
-40/150 |
AHP-1200C31 |
9-35KB-6-C31 |
CCP-31H |
6.0X3.0 |
-76 |
100-240 |
6.8 |
0/50 |
-50/150 |
The current draw of the unit reflect the operation in 120 VAC 60 Hz.
Cold Plate Features:
•Precision machined cold plate surface
•Easy clean stainless steel top surface
•Cools and heats
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Variable fan speed for quieter operation
•Weighs only 25 lbs.(11.4 kg)
•Compact bench top unit,11.2”X 15.1”footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Accessories for glassware (beaker/test tube) cooling
•Many options and accessories available |
Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
Click on the image for a larger view
Click on the image for larger view of the performance curve
|
Other test curves for AHP-1200CAS which might be of interest:
- CCP-22 with AHP-1200CPV CAS, 60 °C to -30 °C cooling rate, CCP-22 was covered
- CCP-22 with AHP-1200CPV CAV, -35 °C to -20 °C step change, CCP-22 was covered
- CCP-31 with AHP-1200CPV CAS range
- CCP-31 with AHP-1200CPV CAS, 1st half insulated and covered, 2nd half covered only
- CCP-31 with AHP-1200CPV CAS response time,
1st half insulated and covered, 2md half covered only
- CCP-31 with AHP-1200CPV CAS response time fan speed @ 50%, covered only
- CCP-31 with AHP-1200CPV CAS delta T, fan speed @ 50%, covered
|
|
|
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view AHP-1200CAS 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
AHP-301CPV Series
Style: Bench Top
Voltages: 100 - 240 VAC, 24 VDC
Cold Plate Temperature Range: -15 °C to 90
°C
Environments: Bench Top, Laboratory, Industrial |
Cooling capacity:
85 Watts @ 0 °C
ΔT
Includes:
Integral Power Supply
Integral PWM Temperature control
USB comms port
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
NOTES |
COLD PLATE |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT °C |
AHP-301CPV |
9-70KB-1-0A0 |
Heat/Cool |
Smooth Surface |
100-240 |
2.0* |
0-45 |
AHP-301CPV |
9-70K5-1-0A0 |
Heat/Cool |
Smooth Surface |
(24) |
7 |
0-45 |
AHP-301CPV |
9-70KB-1-TAP |
Heat/Cool |
6-32 Tap Pattern |
100-240 |
2.0* |
0-45 |
AHP-301CPV |
9-70K5-1-MET |
Heat/Cool |
M3 Tap Pattern |
100-240 |
2.0* |
0-45 |
AHP-301CPV |
9-70K5-1-TAP |
Heat/Cool |
6-32 Tap Pattern |
(24) |
7 |
0-45 |
AHP-301CPV |
9-70K5-1-MET |
Heat/Cool |
M3 Tap Pattern |
(24) |
7 |
0-45 |
* Refelects the current draw of the unit @ 120 VAC, 60 Hz input.
Cold Plate Features:
•Precision machined cold plate surface
•Easy clean stainless steel top surface
•Cools and heats (TYP.-15 °C to 90 °C)
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Weighs only 13 lbs.(5.9 kg)
•Compact bench top unit,9.8”X 10.1”footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Accessories for glassware (beaker/test tube) cooling
•Many options and accessories available |
Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
Click on the image for a larger view
Shown above AHP-301CPV and AHP-301CPV with CH-301 hinged cover accessory.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view AHP-301CPV 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
AHP-301CPV High Temp. Series
Style: Bench Top
Voltages: 100 - 240 VAC, 24 VDC
Cold Plate Temperature Range: -15 °C to 120
°C
Environments: Bench Top, Laboratory, Industrial |
Cooling capacity:
85 Watts @ 0 °C
ΔT
Includes:
Integral Power Supply
Integral PWM Temperature control
USB comms port
Remote Sensibility |
Click on part number to request a quick quote
MODEL NUMBER |
PART NUMBER |
NOTES |
COLD PLATE |
VOLTAGE VAC (VDC) |
CURRENT AMPS. |
OPERATING AMBIENT °C |
AHP-301CPV |
9-70KB-6-0A0 |
Heat/Cool |
Smooth Surface |
100-240 |
2.0* |
0-45 |
AHP-301CPV |
9-70K5-6-0A0 |
Heat/Cool |
Smooth Surface |
(24) |
7 |
0-45 |
AHP-301CPV |
9-70KB-6-TAP |
Heat/Cool |
6-32 Tap Pattern |
100-240 |
2.0* |
0-45 |
AHP-301CPV |
9-70K5-6-MET |
Heat/Cool |
M3 Tap Pattern |
100-240 |
2.0* |
0-45 |
AHP-301CPV |
9-70K5-6-TAP |
Heat/Cool |
6-32 Tap Pattern |
(24) |
7 |
0-45 |
AHP-301CPV |
9-70K5-6-MET |
Heat/Cool |
M3 Tap Pattern |
(24) |
7 |
0-45 |
* Refelects the current draw of the unit @ 120 VAC, 60 Hz input.
Cold Plate Features:
•Precision machined cold plate surface
•Easy clean stainless steel top surface
•Cools and heats (TYP.-15 °C to 90 °C)
•100-240 VAC universal input
•Low-profile design with ergonomic sloped front
•Weighs only 13 lbs.(5.9 kg)
•Compact bench top unit,9.8”X 10.1”footprint
•Virtually maintenance-free operation
•Painted Enameled stainless steel exterior housing
•Accessories for glassware (beaker/test tube) cooling
•Many options and accessories available |
Control Features:
•Integral "tunable" PWM temperature control
•Bi-directional Temperature Control for Thermoelectric applications
•4 Temperature zone with 4 independent PID settings
•Heating and Cooling
•Internal RTD sensor
•Remote Sensibility™ switchable exterior sensor
•Output PWM 1KHz, duty cycle resolution=1000 steps
•Labview VI examples available
•USB communications |
Click on the image for a larger view
Shown above AHP-301CPV and AHP-301CPV with CH-301 hinged cover accessory.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view AHP-301CPV 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
TGB-2515 Thermal Gradient Bar
Gradient Area: 250 X 150 mm
Material: ALuminum, clear anodized
Embeded Sensor: 3 Wire RTD, 100 ohm |
TECA’s TGB-2515 Gradient Bar option expands the use of the AHP-301CPV line of Cold Plates. With a TECA model AHP-301CPV cold plate at each end the gradient bar has programmable gradient areas with near linear gradients between the two ends and near uniform temperatures along the width of the bar. The temperature at each end of the gradient area can be programmed to a single set point or with a ramp/step and soak profile creating constant, expanding, contracting or moving gradients. Set the gradient profile to be large for initial observations then adjust the end set points to zoom in on a temperature range of interest. The temperature at the end of the gradient area is measured from underneath the plate at approximately .05 from the surface leaving the top gradient area clean and smooth. Optional external surface sensors can be used to change the size of the gradient area. |
Click on the image for larger view
Picture shows TGB-2515 Thermal Gradient Bar used with two AHP-301CPV
Click on the image for a larger view of the performance curve
CLick on the image for larger view
Click on the image to open the PDF file
|
|
|
AHP-800MSP Magnetic Stirring Plate
Style: Bench Top
Voltages: 100-240 VAC
Environments: Bench Top, Laboratory, Industrial |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Integral Power Supply
Integral Temperature control
RS-232 Communications
Remote Sensibility
Stock cold shoe for 4 Liter Corning Pyrex Brand 1000 Beakers |
Click on part number to request a quick quote
MODEL NUMBER
|
PART NUMBER |
NOTES |
VOLTAGE VAC |
CURRENT AMPS. |
TEMP.
CONTROL |
AMBIENT RANGE °C |
AHP-800MSP |
9-50EB-1-003 |
Heat/Cool |
100-240 |
4.0 |
TC-4300 |
0/45 |
Features:
• Heating and cooling
• Unique high-performance side mounting cold plate for added cooling and uniform temperatures
• Magnetic drive for stir bars from underneath the bottle
• Magnetic stir offers 5 speed settings
• Made for use with Corning Pyrex brand 1000 beakers
• 100-240 VAC universal, Integral power supply
• Height adjusting rubber feet
• Power input cord set
• Weighs only 38 lbs. (17.3 kg)
• Compact bench-top design
• No compressor, fluorocarbons or filters
• Virtually maintenance-free operation
• Painted stainless steel exterior housing
• Removable bottle block for different size bottles or beakers |
Includes:
• Integral PID "tunable" temperature control
• One shot smart PID control tuning or Adaptive Smart Continuous Tuning
• Internal RTD sensor
• Remote Sensibility™ switchable exterior RTD sensor
• Multi-segment ramp and soak programs
• RS-232 communications
• i-tools software for easy programming and controltweaking
• Optional software for charting and data acquisition
• Stock bottle block for 4 liter beaker (Corning Pyrex brand
1000 beaker)
|
Optional Sleeves:
The AHP-800MSP has optional sleeves that make it adaptable to different size bottles and beakers. Install different sleeves for different diameter vessels.
SLV-2000 for 2.0 Liter beaker
SLV-1500 for 1.5 Liter beaker
SLV-1000 for 1.0
Liter beaker
SLV-800 for 0.80 Liter beaker
SLV-600 for 0.60 Liter beaker |
Click on the image for a larger view
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view AHP-800MSP 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
CCP-22 Cascade Cold Plate
Description: Low Temperature Cascade
Voltages: 24 VDC
Cascade Temperature Range: -30 °C to 90
°C
Requirement:
AHP-1200CPV for Cascades Part Number: 9-35EB-1-CAS |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Internal RTD Sensor |
Click on the image for a larger view
Shown above CCP-22 cascade and CCP-22 cascade mounted on AHP-1200CPV CAS.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view CCP-22 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
CCP-22H High Temp. Cascade Cold Plate
Description: Extended Temperature Cascade
Voltages: 24 VDC
Cascade Temperature Range: -30 °C to 150
°C
Requirement:
AHP-1200CPV for Cascades Part Number: 9-35EB-6-CAS |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Internal RTD Sensor |
Click on the image for a larger view
Shown above CCP-22 cascade and CCP-22 cascade mounted on AHP-1200CPV CAS.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view CCP-22 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
CCP-21 Cascade Cold Plate
Description: Low Temperature Cascade
Voltages: 24 VDC
Cascade Temperature Range: -40 °C to 90
°C
Requirement:
AHP-1200CPV for Cascades Part Number: 9-35EB-1-CAS |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Internal RTD Sensor |
Click on the image for a larger view
Shown above CCP-21 low termperature cascade and CCP-21 mounted on AHP-1200CPV CAS.
Click for larger view of the performance curve
|
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view CCP-21 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
CCP-21H High Temp. Cascade Cold Plate
Description: Extended Temperature Cascade
Voltages: 24 VDC
Cascade Temperature Range: -40 °C to 150
°C
Requirement:
AHP-1200CPV for Cascades Part Number: 9-35EB-6-CAS |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Internal RTD Sensor |
Click on the image for a larger view
Shown above CCP-21 low termperature cascade and CCP-21 mounted on AHP-1200CPV CAS.
Click for larger view of the performance curve
|
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view CCP-21 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
CCP-31 Cascade Cold Plate
Description: Low Temperature Cascade
Voltages: 24 VDC & 12 VDC
Cascade Temperature Range: -50 °C to 90
°C
Requirement:
AHP-1200CPV for Cascades Part Number: 9-35EB-1-CAS |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Internal RTD Sensor |
Click on the image for a larger view
Shown above CCP-31 low termperature cascade and CCP-31 mounted on AHP-1200CPV CAS with CH-1200 hinged cover accessory.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view CCP-31 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
CCP-31H High Temp. Cascade Cold Plate
Description: Extended Temperature Cascade
Voltages: 24 VDC & 12 VDC
Cascade Temperature Range: -50 °C to 150
°C
Requirement:
AHP-1200CPV for Cascades Part Number: 9-35EB-6-CAS |
Applications:
Laboratory or industrial environments, drugs and industrial chemicals. Process testing. Quality control.
Includes:
Internal RTD Sensor |
Click on the image for a larger view
Shown above CCP-31 low termperature cascade and CCP-31 mounted on AHP-1200CPV CAS with CH-1200 hinged cover accessory.
Click on the image for a larger view
View or request 3D solid model files for this product:
Click on the image to view CCP-31 3D PDF file: |
|
Click on the image to open the PDF file
|
|
|
TECA's thermoelectric (Peltier) Air-Cooled Laboratory Cold/Hot Plates are perfect for use in laboratories or anywhere solid-state cooling or heating surfaces are needed. Applications include cooling and heating lab samples, tissue dissection (histology), quality control, pharmaceuticals, research, pain studies (thermal nociception) and more. Notably, the AHP-301CPV and AHP-1200DCP models have been used in published research articles (see 50 Years of Innovation section on this website for details, 2005 and 2009-2011).
Each thermoelectric Cold/Hot plate on this page include versatile features such as: integral programmable temperature controller, data logging software, RS-232 communications, integral RTD sensor, ability to switch to exterior sensor, and more.
Cold/Hot Plate surfaces can be smooth or include tapped accessory holes at no extra charge (see drawing). Accessories are available to help you further customize the plate to your application.
Almost every model has a high temperature version available for applications requiring heating above 90C.
Single Cold/Hot Plates include the thermoelectric AHP-301CPV and AHP-1200CPV. We also offer a Dual Cold/Hot Plate, the thermoelectric AHP-1200DCP. This unique Peltier product is intended primarily for pain research and behavior studies using lab animals. However, this product has been customized for other uses including large sample cooling and heating. Also of note, Cascade Cold Plate systems allow for cooling to temperatures as low as -50C. |
|
|
|
|
|
|